16.2 Specifications
Guidance and manufacturer’s declaration - electromagnetic emissions
All values stated are nominal.
Height:
Depth:
Width:
Weight:
The A-V Impulse 6000 Series Controller is intended for use in the electromagnetic environment specified below.
The customer or the user of the A-V Impulse 6000 Series Controller should assure that it is used in such an environment.
16.5 cm overall
18.6 cm
32 cm
3.57 kg
Electrical Supply: 120V 0.3A 60Hz
Fuses: T1AH 250V x 20 mm (Antisurge)
Standards: EN/IEC 60601-1, EN/IEC 60601-1-2, UL 60601-1, CAN/CSA C22.2 No. 601.1-M90, FCC Part 15
Class 1 Type B device
Output Impulse Pressure: 60 - 200 mmHg
Impulse Duration: 1 second / 3 second (3 second factory set default)
Impulse Interval: 20 - 50 seconds (20 seconds factory set default) Operating conditions: Temperature 0°C - +45°C Transportation and storage conditions: Temperature -40°C to +70°C
Relative humidity: 10%-95% condensing Atmospheric pressure: 500mbar-1060mbar Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The A-V Impulse 6000 Series Controller uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The A-V Impulse 6000 Series Controller is suitable for use in all establishments,
including domestic establishments and those directly connected to the public lowvoltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The A-V Impulse 6000 Series Controller is intended for use in the electromagnet environment specified below. The customer or the
user of the A-V Impulse 6000 Series Controller should assure that it is used in such an environment.
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Electrical fast transient/
burst IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply lines
±1 kV for input/output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines IEC 61000-4-11
<5 % UT (>95 % dip in UT)
for 0.5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT)
for 5 sec
<5 % UT (>95 % dip in UT)
for 0.5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT)
for 5 sec
Mains power quality should be that of a typical
commercial or hospital environment. If the
user of the A-V Impulse 6000 Series Controller
requires continued operation during power
mains interruptions, it is recommended that
the A-V Impulse 6000 Series Controller be
powered from an uninterruptible power supply
or a battery.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
16.3 Impulse Pressure Conversion Table mmHg/KPa
mmHG
KPa
mmHG
KPa
mmHG
KPa
60
70
80
90
100
8.0
9.3
10.7
12.0
13.3
110
120
130
140
150
14.7
16.0
17.3
18.7
20.0
160
170
180
190
200
21.3
22.7
24.0
25.3
26.7
17. Guidance and Manufacturer’s Declaration - Electromagnetic Emission
Warning: Medical electrical equipment needs special precautions regarding EMC and needs to be
installed according to the EMC information provided. Careful consideration of this information is
essential when stacking or collocating equipment and when routing cables and accessories.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Warning: RF mobile communications equipment can effect medical electrical equipment.